Vaisala GMP252 - Spare Parts and Accessories

Vaisala GMP252
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Property Description/Value
with compensation, 0 ... 10 000 ppm%CO
2
,
0 ... 90 %O
2
±0.6 % of reading (at +25 °C (+77 °F))
without compensation (typical) -0.08 % of reading/%O
2
Start-Up, Warm-Up and Response Time
Start-up time at +25 °C < 12 s
Warm-up time for full spec. < 2 min
Response time (T90) with standard filter < 1 min
Response time (T90) with spray shield < 3 min
Flow-Through Option
Response time (T90) with > 0.1 l/min 30 s
Flow rate dependence < 1 l/min flow no eect
Flow rate dependence 1 ... 10 l/min flow < 0.6% of reading l/min
Gas flow operating range < 10 l/min
Gas flow recommended range 0.1 ... 0.8 l/min
Table 48 Operating Environment
Property Description/Value
Operating temperature of CO
2
measurement -40 ... +60 °C (-40 … +140 °F)
Storage temperature -40 ... +70 °C (-40 … +158 °F)
Humidity 0 ... 100 %RH, non-condensing
Condensation prevention Sensor head heating when power on
EMC compliance EN61326-1, Generic Environment
Chemical tolerance (temporary exposure during
cleaning)
H
2
O
2
(2000 ppm, non-condensing)
Alcohol-based cleaning agents (for example
ethanol and IPA)
Acetone
Acetic acid
Pressure
Compensated 500 ... 1100 hPa
Operating < 1.5 bar
Table 49 Mechanical Specifications
Property Description/Value
Weight, probe 58 g (2.05 oz)
GMP252 User Guide M211897EN-D
104