Calibration and Adjustment (page 49)
Configuration Registers (page 114)
Calibration and Adjustment with MI70 Indicator (page 89)
Indigo Calibration Overview (page 78)
9.3.5 DRW244827SP Calibration Adapter
The optional calibration adapter accessory can be used to feed a reference gas to the probe
through a gas port when calibrating. Gas flow should be in the range 0.5 ... 1.1 l/min,
recommendation is 1 l/min. Allow the measurement to stabilize for three minutes before
starting the calibration.
1
3
2
Figure 32 DRW244827SP Calibration Adapter with Probe Inserted
1 O-ring inside the adapter
2 Gas outlet on each side of the adapter
3 Gas port (port outer diameter 4.6 mm, port hole inner diameter 2 mm, suitable for
tubing with 4 mm inner diameter)
Chapter 9 – Maintenance
97