Table 39 Pcmode Command
Syntax Description
pcmode<cr> Check current pressure compensation mode.
Possible modes:
on = Compensation enabled using setpoint value.
off = Compensation disabled, default (neutral)
value used: see Environmental Compensation
(page 15).
pcmode [on | off]<cr>
Change pressure compensation mode (on or off).
Example (check pressure compensation mode; pressure compensation is enabled using a setpoint value):
pass 1300
pcmode
P COMP MODE : ON
Table 40 Rhcmode Command
Syntax Description
rhcmode<cr>
Check current relative humidity compensation mode.
Possible modes:
on = Compensation enabled using setpoint value.
off = Compensation disabled, default (neutral)
value used: see Environmental Compensation
(page 15).
To apply an accurate relative
humidity compensation, make
sure that the temperature
compensation and pressure
compensation configurations also
match the measurement
environment.
rhcmode [on | off]<cr>
Change relative humidity compensation mode (on or
off).
Example (check relative humidity compensation mode; relative humidity compensation is disabled, a neutral
value is used):
pass 1300
rhcmode
RH COMP MODE : OFF
GMP252 User Guide M211897EN-D