Syntax Description
Example (for channel 1, show the currently set analog output parameter and scaling):
pass 1300
asel 1
Aout 1 quantity : CO2(0 ... 10000 ppm)
Example (for channel 1, set scaling to 0 ... 4000 ppmCO
2
):
pass 1300
asel 1 co2 0 4000
Aout 1 quantity : CO2(0 ... 4000 ppm)
4.12 Calibration and Adjustment
Before using the calibration and adjustment commands, read through
Calibration and Adjustment (page 94). Make sure that the environmental
compensation settings of the probe are properly set for your calibration environment;
see Environmental Compensation (page 15).
CAUTION!
Table 32 Adate Command
Syntax Description
adate<cr>
Show CO
2
factory adjustment date.
Example:
adate
Adjustment date : 20150420
Table 33 Atext Command
Syntax Description
atext<cr> Show CO
2
factory adjustment information.
Example:
atext
Adjusted at Vaisala/Helsinki
Chapter 4 – Vaisala Industrial Protocol
49