Table 34 Cdate Command
Syntax Description
cdate<cr>
Show calibration date.
cdate [yyyymmdd]<cr>
Set a new calibration date.
yyyymmdd = Year (yyyy), month (mm) and day (dd)
of calibration
Example:
pass 1300
cdate
Calibration date : 20150220
Example (set a new calibration date to June 30, 2015):
cdate 20150630
Calibration date : 20150630
Table 35 Ctext Command
Syntax Description
ctext<cr>
Show calibration information text.
ctext [text]<cr>
Set a new calibration information text to be shown
after the automatic text "Calibrated at".
Example:
pass 1300
ctext
Calibrated at 500 ppm in lab
Example (set a new information text):
ctext 0/1000 by NN
Calibrated at 0/1000 ppm by NN
Table 36 CCO2 Command
Syntax Description
cco2<cr>
Show current user adjustment status.
GMP252 User Guide M211897EN-D