Syntax Description
Example (target probe in POLL mode, with address 52):
open 52
GMP25x: 52 Opened for operator commands
Table 23 R Command
Syntax Description
r<cr>
Start the continuous outputting of measurement
values as an ASCII text string to the serial line. The
probe keeps outputting measurement messages at
the interval that has been set with the intv
command until stopped with the s command.
Example:
r
CO2= 1024 ppm
CO2= 1024 ppm
CO2= 1028 ppm
CO2= 1026 ppm
CO2= 1028 ppm
...
Table 24 S Command
Syntax Description
s<cr>
Stop the continuous outputting that was started with
the r command.
Example:
...
CO2= 658 ppm
CO2= 654 ppm
CO2= 655 ppm
s
Chapter 4 – Vaisala Industrial Protocol
41