Syntax Description
Example (enable temperature, pressure and relative humidity compensation using setpoint values):
pass 1300
tcmode on
T COMP MODE : ON
pcmode on
P COMP MODE : ON
rhcmode on
RH COMP MODE : ON
Table 41 Tcmode Command
Syntax Description
tcmode<cr>
Check current temperature compensation mode.
Possible modes:
on = Compensation enabled using setpoint value.
off = Compensation disabled, default (neutral)
value used: see Environmental Compensation
(page 15)
measured = Compensation enabled using internal
measurement.
tcmode [on | off | measured]<cr>
Change temperature compensation mode (on, off or
measured).
Example (check temperature compensation mode; temperature compensation is enabled using a setpoint
value):
pass 1300
tcmode
T COMP MODE : ON
Example (change temperature compensation to use internal measurement):
pass 1300
tcmode measured
T COMP MODE : MEASURED
Chapter 4 – Vaisala Industrial Protocol
57