2. Set the serial mode to Modbus with the smode command:
smode modbus
3. Set the Modbus address to 240 with the addr command:
addr 240
4. Set the serial line settings to 19200/N/8/2 with the seri command:
seri 19200 N 8 2
5. Power o (disconnect) the probe or reset with the resetcommand. The new
configuration is available at the next restart.
More Information
Accessing Serial Commands from Modbus or Analog Mode (page 30)
4.7
 Changing From Digital Output to
Analog Output
1. Set up a terminal connection as instructed in Connecting with a Computer (page 27).
2. Change the mode from digital to analog with the smode serial command: smode
analog.
3. Reset the probe (disconnect and reconnect the cable or use the reset serial
command) to power on in analog output mode.
More Information
Accessing Serial Commands from Modbus or Analog Mode (page 30)
4.8
 Serial Commands
The notation <cr> refers to the carriage return control character, which you can send in a
terminal application by pressing enter on your keyboard. Before entering commands, send a
<cr> to clear the command
buer.
You can enter the commands in uppercase or lowercase. In the command examples, the
keyboard input by the user is in bold type.
Chapter 4 – Vaisala Industrial Protocol