Figure 29 CO
2
Reading with Tcomp and Pcomp on MI70 Screen
You can change the compensation settings from Main menu > Settings > Measurement
settings.
Figure 30 Probe Compensation Settings on MI70 Screen
When you turn a compensation
o, the probe still shows a value for the corresponding
display quantity (for example, Pcomp shows 1013.2 hPa). This is the default compensation
value that is mathematically neutral for the probe’s internal compensation model.
GMP252 cannot receive the compensation values from a second probe connected to MI70.
Only direct configuration is supported.
8.10 Calibration and Adjustment with
MI70 Indicator
Before using the MI70 indicator for calibration and adjustment, read the instructions in
Calibration and Adjustment (page 94). Make sure that the environmental compensation
settings of the probe are properly set for your calibration environment; see Changing
Environmental Compensation Settings with MI70 Indicator (page 88).
Chapter 8 – Operating with MI70 Indicator
89