Table 37 Env Command
Syntax Description
env<cr>
Show current compensation values.
Before using this command, you must enable
environmental compensation using the following
commands:
• o2cmode [on]
• pcmode [on]
• rhcmode [on]
• tcmode [on | measured]
env [temp | pres | oxy | hum]
[value]<cr>
Set new permanent compensation values and store
them in eeprom.
Eeprom:
• Non-volatile memory, values retained during power
o.
• Number of writes is limited to 30000 cycles by
memory implementation.
• Must only be used for writing permanent values, to
avoid wearing out the eeprom.
temp = Compensation temperature. Range -40 ...
+100 °C.
pres = Compensation pressure. Range 500 ... 1100
hPa.
oxy = Oxygen content of background gas. Range 0 ...
100 %.
hum = Relative humidity of background gas. Range
0 ... 100 %.