Table 25 Sdelay Command
Syntax Description
sdelay<cr>
Show serial line transmission delay (1 … 255). 1 unit = 4
milliseconds (for example, 5 = 0.020 second
transmission delay).
sdelay [delay]<cr>
Set a new serial line transmission delay.
delay = Serial line delay, range 1 … 255 (1 unit = 4
ms).
Example (set delay to 0.1 seconds = 25 units):
sdelay 25
COM transmit delay : 25
Table 26 Send Command
Syntax Description
send<cr>
Output a single measurement message.
send [aaa]<cr>
Output a single measurement message from a device
in poll mode.
aaa = address of the probe, range range 0 ... 254
Example:
send
CO2= 1422 ppm
Example (target probe in POLL mode, with address 52):
send 52
CO2= 458 ppm
Table 27 Seri Command
Syntax Description
seri<cr>
Show current serial line settings.
GMP252 User Guide M211897EN-D