By default, the filtering factor is set to 1.0 (1.0 = 100 in the configuration range), which means
the latest measurement is shown directly in the output, without any filtering. To apply
filtering, enter a lower filtering factor to include previous measurements in the calculation of
measurement output. For example, changing the filtering factor to 0.1 results in an output
that is a combination of previous measurements (90%) and the latest measurement (10%).
The configuration range of the filtering factor is 0 ... 100: for example, to set the factor to 0.5,
set the value to 50.
More Information
‣
Filtering Factor (page 120)
7.12 Calibrating GMP252 with Indigo
7.12.1 Indigo Calibration Overview
You can carry out 1-point and 2-point adjustments with the Indigo transmitter's wireless
configuration interface. In addition to calibrating and adjusting probes, you can view the
current adjustments and restore the probe's factory adjustment.
Figure 25 Calibration Menu Main View
1 Calibration tab
2 Configuration tab
3 Diagnostics tab
4 Measurements tab
There are 4 tabs in the Calibration menu:
• Calibration: the main adjustment view with options for making adjustments, viewing
adjustments, and restoring factory adjustments.