Syntax Description
aover [channel clipping error_
limit]<cr>
Set the behavior of the analog output when the
measured value is outside the scaled output range.
channel = Analog output channel
1 = voltage output (V)
2 = current output (mA)
clipping = Output margin (%) at which the output
is clipped.
error_limit = Measurement value margin (%) at
which the output of the channel goes into the error
state. The current or voltage output of the error state
is defined using the amode command.
Example (view currently set analog output overrange behavior on channel 1):
pass 1300
aover 1
Aout 1 clipping :5.00 %
Aout 1 error limit :10.00 %
GMP252 User Guide M211897EN-D