Responding to System Alarms
6-12 Force FX-8C Service Manual
81 Internal component malfunction.
Optoisolator test failed
Do not attempt to use the generator. Record the number and call
the Valleylab Service Center.
90
95
Generator model number in main
microcontroller ROM not Force FX-8C
model number
Generator model number in main
microcontroller ROM not Force FX-8C
model number
Refer to Correcting IC U6 Malfunctions (steps 1–7) and Correcting
IC U3 Malfunctions (steps 4–11) in this section.
100
101
102
103
104
105
110
111
Software malfunction. rem_update_state
value outside state range
Software malfunction. rem_pad_state
value outside state range
Software malfunction. rem_state value in
re_rem_control( ) outside state range
Software malfunction. rem_state value in
re_non_rem_control( ) outside state range
Software malfunction. rem_flash_state
value outside state range
Software malfunction. rem_led_state value
outside state range
Software malfunction. cal_state value in
ca_generator_setup( ) outside state range
Software malfunction. cal_state value in
ca_clock_date_setup( ) outside state
range
Turn off, then turn on the generator. If the alarm number
reappears, record the number and call the Valleylab Service
Center.
Number Description Recommended Action

Table of Contents