Multi-View
VentecLife.com 105
Reading Multi-View Reports
This section provides detailed information on using and reading VOCSN Multi-View reports. Each section of the report
provides insight into aspects of V*Home therapy use and monitored trends.
Report Information
The rst page of VOCSN Multi-View reports contain
information about the V*Home device, as well as
the report itself. This page includes the device
serial number, its conguration and the software
version installed at the time of the Multi-View
export. It also includes information about the report
creation date, duration, and the date and time
range of data included.
NOTE: To maintain data continuity, the dates
and times used to display Multi-View data are
anchored to the most recent V*Home Date and
Time control settings. Unlike the Event Log
displayed on the V*Home screen (which stamps
events with whatever the V*Home Date and Time
settings were at the time the event occurred),
Multi‑View time is calculated backwards so
that event times are relative to the most recent
setting. This approach ensures the continuity of
graphed and trended data in Multi‑View reports,
even through time zone, daylight savings, or
other changes to the Date and Time controls on
V*Home.
The report cover page also includes user-editable
elds that can be used to record additional
information about the report. To use these elds, open the PDF le, click one of the blank line items, and type in the
desired value. Saving the report will also save the date you entered.
NOTE: If you need to ensure that the data you enter into the form elds cannot be edited again, use a PDF reader to ll
in the elds, and then select “Print” to print the le to PDF and lock your entries in place in the new le.