Controls and Settings
VentecLife.com 75
Changing Nebulizer Compensation Settings
Nebulizer therapy can congured to
compensate for the ow added to
the patient circuit from an external
nebulizer. The Nebulizer Duration
control may be set for a period of
time between 5 and 60 minutes. To
run Nebulizer therapy, rst set the
Nebulizer Duration control:
1 Press the Therapy tab, and
then press the NEBULIZER
button.
2 Press the plus (+) and minus
(-) buttons on the right side
of the screen to congure the
Nebulizer Duration.
3 While Nebulizer therapy is
running, you may use the plus (+) button to add more time if needed.
4 Press the EXIT button to return to the main menu. The next time you access the Nebulizer therapy screen, the
Nebulizer Duration setting will remain at whatever it was last set to.
The following Nebulizer therapy control is available on V*Home.
NOTE: Bold black text indicates the default setting.
Nebulizer Control Settings Description
Nebulizer
Duration
5 to 60
minutes
6 minutes
The Nebulizer Duration control sets the duration over which V*Home drives a connected
nebulizer, or compensates for the ow from an external one.
The resolution of the Nebulizer Duration control is 1 minute.
0 5 10 15 20 25 30L H
MINUTES
NEBULIZER DURATION
EXIT
60
5
10
15
20
55
50
45
40
2535
30
15
mins
No Alarms
PRESSURE (cmH20)
SIMV-VOL
V N
START
EXTERNAL NEB.
COMPENSATION
Press here to
EXIT this
Press here to
increase or
decrease how
long the
nebulizer runs
Press here to select
external nebulizer
compensation