Controls and Settings
VentecLife.com 61
Available Ventilation Therapy Settings
The following table lists the congurable Ventilation therapy controls available on V*Home. These controls are available
separately for each of the three congurable Ventilation therapy Presets.
NOTE: Controls that cannot be edited will appear in gray. Controls may be locked because of the Permissions and
Clinician Unlock feature. In addition, the Mode and Circuit Type controls will appear in gray and cannot be edited while
the selected Ventilation Preset is running. These three controls can only be modied in an inactive Preset.
NOTE: The resolution of each control with numerical settings is 1, unless otherwise stated in the control description.
NOTE: Bold black text indicates the default setting.
Ventilator
Control
Settings Description
Apnea Rate
4 to 60 BPM Backup
12 BPM Backup
The Apnea Rate control sets the breath rate at which mandatory breaths will be
delivered to the patient while the Apnea alarm is activated.
NOTE: The Apnea Rate control is not available when Circuit Type is set to
Mouthpiece.
Breath Rate
0 to 60 BPM
12 BPM
The Breath Rate control congures the minimum number of breaths per minute
(BPM) delivered to the patient.