Controls and Settings
VentecLife.com 76
Clinician Access Mode
When the Unlock Required? device setting is set to Yes, pressing the CLINICIAN ACCESS
button locks and unlocks V*Home. Using the Permissions feature, V*Home can be congured
to restrict which controls are congurable by all users, and which control settings can only be
changed by rst entering a Clinician Access Passcode to unlock V*Home.
When V*Home is unlocked, all therapies and device settings are congurable. Unlocking
V*Home using the Clinician Access Passcode also allows access to the Permissions feature,
where each control can be set to allow only clinician access (by rst using the passcode), or
set to allow conguration by all users at all times. See “Conguring Permissions” on page
77 for more information.
Once the Clinician Access Passcode is entered, V*Home will remain unlocked for 15 minutes. V*Home will automatically
lock again after 15 minutes, regardless of user interaction with the device. V*Home can also be locked again at any time
by pressing the Clinician Access button in the Menu tab.
Entering the Clinician Access Passcode
To enter the Clinician Access Passcode and unlock V*Home:
1 Press the Menu tab.
2 Press the CLINICIAN ACCESS
button.
3 Enter the four-digit Clinician
Access Passcode. The Clinician
Access Passcode is the last
four digits of the device serial
number. This number is printed
on the V*Home back label,
and is also visible from the MY
VOCSN screen.
4 Press OK. V*Home will unlock,
allowing the conguration of all
therapy controls, permissions,
and device settings.
NOTE: When the Unlock Required? device setting is set to Yes, V*Home will automatically lock after 15 minutes. To lock
V*Home manually, press the CLINICIAN ACCESS button on the Menu screen.
CLINICIAN
ACCESS
0 5 10 15 20 25 30L H
Menu
MY VOCSN
i
LOGSPRE-USE TEST DEVICE
SETTINGS
Therapy MenuMonitors
Home
SETTINGS
& ALARMS
CLINICIAN
ACCESS
Click OK to enter Clinician
Access mode.
Please enter passcode
for clinician access
ACCEPTCANCEL
1 2 3
4 5 6
7
0
8 9
No Alarms
PRESSURE (cmH20)
SIMV-PRES
V

Table of Contents