Technical Specications
VentecLife.com 158
Monitor Resolution and Accuracy
NOTE: When used with passive, or valveless Ventec One‑Circuits, V*Home was designed for use with a humidier or
HME. All volumes and ows are expressed in BTPS unless stated otherwise.
NOTE: Monitors incorporating pressure measurements are calculated using a pressure sensor inside V*Home, which
calculates the pressure at the Ventec One-Circuit connection port.
NOTE: Unless otherwise stated, monitor accuracy specications are met when V*Home is used as recommended by
Ventec Life Systems.
Monitor Resolution Accuracy
Airway pressure
manometer
N/A ±(1.4 cmH2O + 8% of actual) up to 15 cmH2O
±(2 cmH2O + 4% of actual) above 15 cmH2O
Breath Rate 1 BPM ±10% BPM
Date 1 day N/A
FiO2 1% ±(2.5% + 2.5% of actual)
NOTE: Air contaminants and pollutants (including methane) may have an adverse
effect on the accuracy and long‑term performance of the FiO2 monitor.
I:E Ratio 0.1 Calculated from the average monitored inspiratory time and average monitored
exhalation time accuracies of ±50 ms or 5%, whichever is greater
Internal Battery Capacity N/A N/A
Leak 1 L/min ±(10% + 1 L/min)
MAP 1 cmH2O See Airway pressure manometer
Minute Volume 0.1 L when
≤9.9
1.0 L when
≥10
Calculated (for active circuits) or estimated (for passive and valveless circuits) from the
average exhaled tidal volume (Vte) and number of breaths per minute (BPM) for the last
8 breaths
Nebulizer Minutes
Remaining
1 minute ±1 minute
Patient Triggered N/A N/A
PEEP 1 cmH2O ±(1.4 cmH2O + 8% of actual) up to 15 cmH2O
±(2 cmH2O + 4% of actual) above 15 cmH2O
PIP (Peak Inspiratory
Pressure)
1 cmH2O ±(1.4 cmH2O + 8% of actual) up to 15 cmH2O
±(2 cmH2O + 4% of actual) above 15 cmH2O