Alarms
VentecLife.com 93
Clearing an Alarm
Visual indicators of an alarm condition remain on the screen as a count in the status bar and a list in the Alarm Log until
the alarm condition is resolved and the list is cleared. To clear an activated alarm:
1 Press the Alarm Silence button on the front of V*Home to silence the alarm tones, if desired.
2 Resolve the alarm condition by taking the appropriate action. When an alarm condition is resolved, the alarm will
clear. A log of the alarm activation will be recorded in the Alarm Log and the Event Log.
3 Access the Alarm Log through the Menu tab, or by pressing the Alarm banner in the upper left corner of the
V*Home touchscreen, then select CLEAR LIST to remove all but active alarms from the Alarm Log.
NOTE: A record of recent alarm conditions will remain in the Event Log, even when the Alarm Log has been cleared.
4 Press the < EXIT tab to exit the Alarm Log.
Remote Alarms and Nurse Call Systems
(Optional)
Connect an optional remote alarm or nurse call system using the port on the back of V*Home. See “Connecting a Nurse
Call System or Remote Alarm” on page 33 for setup information.
NOTE: The delay between an alarm condition and activation of the remote alarm port is <1 s. Refer to the remote alarm
manufacturer’s instructions for use to determine the maximum possible delay between V*Home alarm and remote alarm
activation.