Changing Between Congured Ventilation Presets
V*Home can be congured with up to three different Ventilation therapy Presets. To begin Ventilation therapy using a
different Preset:
1 Press the Therapy tab, and then press the Ventilation therapy icon.
2 Press the name of the Preset you want to activate.
3 Press START on the bottom of the screen to begin Ventilation therapy using the selected Preset control and
alarm settings.
NOTE: If the Circuit Type control of the selected ventilation Preset is set to a value different than the existing Preset, you
will be prompted to modify the Ventec One‑Circuit accordingly and run a Pre‑Use Test prior to reinitializing Ventilation
therapy. If time constraints make running the test inadvisable, press the EXIT button to immediately initiate Ventilation
therapy with the selected Preset.
NOTE: To change Ventilation therapy settings press the Preset name and then the SETTINGS & ALARMS button. Follow
the instructions described in “Changing Ventilation Therapy Settings” on page 59 to modify Ventilation therapy
settings. When conguration is complete, press < EXIT to return to the Ventilation therapy screen.
0 5 10 15 20 25 30L H
EXIT
SETTINGS
& ALARMS
Prepare patient circuit for ventilation,
select preset, and press START to begin.
AC-PRES
NIGHTDAY
SIMV-VOL
MOBILE
SIMV-PRES
ACTIVE NOWSTARTSTART
No Alarms
PRESSURE (cmH20)
SIMV-VOL
V
NOW ACTIVE