Setup
VentecLife.com 16
Setup
This chapter includes instructions for placing and setting up V*Home, including instructions for connecting Ventec One-
Circuits, and additional Ventec One-Circuit components. V*Home was designed for use with Ventec One-Circuits from
Ventec Life Systems. Do not use third-party patient circuits with V*Home.
WARNING: Put V*Home into service in accordance with the information provided in this Clinical and Technical Manual.
V*Home operation may be impaired or become unsafe by failure to follow setup and operating instructions, the
connection of unauthorized accessories, or the unauthorized modication of V*Home. All modications made, and
accessories used with V*Home, must meet the requirements of IEC 60601-1. The organization responsible for device
setup must ensure the compatibility of V*Home and all parts and accessories used to provide therapy to the patient
prior to use.
WARNING: V*Home should be set up, congured, and used by trained clinicians and caregivers under the direction
of a physician. Patients and caregivers should be instructed not to modify the V*Home setup or conguration without
direction and/or supervision from a clinician.
WARNING: Do not use lubricants on V*Home or on any part of the Ventec One-Circuit. Lubricating ttings,
connections, tubing, or other accessories may result in re and burns to the patient or caregiver.
WARNING: Use only spare parts recommended by Ventec Life Systems. Using spare parts not recommended by the
V*Home manufacturer may result in re and burns to the patient or caregiver.
WARNING: Only use carrying cases approved by Ventec Life Systems. Use of unauthorized carrying cases may result in
damage to V*Home, impaired device performance, and risk to the patient.
NOTE: The V*Home Ethernet and USB ports are intended for use by trained personnel only. The remote alarm port and
power connection port can be used with third‑party equipment. However, it is the responsibility of the person connecting
the third‑party equipment to ensure the system complies with clause 16 of IEC 60601‑1, 3rd ed., as well as any local laws.
NOTE: When using V*Home to provide non‑invasive ventilation, Ventec Life Systems recommends use of CO2 monitoring
equipment compliant with ISO 80601‑2‑55. To connect and use CO2 monitoring with V*Home, follow all instructions from
the equipment manufacturer.