Breath Types and Therapy Modes
VentecLife.com 37
Pressure Breaths
Mandatory and assist breaths are delivered as pressure breaths in pressure ventilation modes. V*Home delivers pressure
breaths by elevating the pressure of the Ventec One-Circuit to the set Pressure Control limit for the set Inspiratory Time.
NOTE: The maximum pressure delivered is limited to the High Pressure alarm setting minus 3 cmH2O.
V*Home will deliver ow during the set Inspiratory Time to reach and maintain the set Pressure Control limit. Adjusting
the Rise Time control will modify the rate of ow and rate of pressure elevation.
Pressure breaths cycle at the end of the set Inspiratory Time, or when ow drops to the set Flow Cycle percentage of
peak ow when PC Flow Termination is set to On.
Set
Pressure
Control
Flow
Breath cycles at set
Inspiratory Time
Default settings with
Rise Time set to 6 and
Pressure Control Flow
Termination enabled
Set
Inspiratory
Time
Set
Inspiratory
Time
Set
Inspiratory
Time
Airway
Pressure
Breath cycles at
set Flow Cycle
Default settings
Default settings with
Rise Time set to 6