Controls and Settings
VentecLife.com 68
Controls Available in Each Ventilation Mode
The table below lists the controls available for conī·guration depending on the set Ventilation Mode. Unavailable controls will not be visible when setting
Ventilation therapy controls.
NOTE: Controls available regardless of the set Ventilation Mode are not listed in the following table.
Control Bi-Level Spontaneous
Assist/
Control-
Pressure
SIMV-
Pressure
Assist/
Control-
Volume
SIMV-
Volume
Vol, Targeted-
PS
Vol-Targeted-
PC
Vol. Targeted-
SIMV High Flow
Apnea Rate
Yes, when
Circuit
Type is
set to
Passive or
Valveless
No Yes, when Circuit Type is set to Passive or Valveless No
Breath Rate
No No
Yes, when
Circuit Type
is set to
Passive or
Valveless
Yes
Yes, when
Circuit
Type is set
toPassive
or
Valveless
Yes No
Yes, when
Circuit Type
is set to
Passive or
Valveless
Yes No
EPAP
Yes No No No No No No No No No
Flow
No No No
Yes, when
High Flow is
set to On
No
Yes, when
High Flow is
set to On
No No
Yes, when
High Flow is
set to On
Yes
Flow
Cycle
Yes Yes
Yes, when
PC Flow
Termination
is set to On
Yes No Yes Yes
Yes, when
PC Flow
Termination
is set to On
Yes No