Technical Description
VentecLife.com 172
Technical Description
Theory of Operation
The sections that follow detail the theory of operation of each of the ve V*Home therapies. A pneumatic schematic of
each therapy is also provided.
Ventilation Therapy
Gas is drawn from the air intake, ltered for dust and particles, and ows into the accumulator. The air is mixed with
oxygen if V*Home is congured to deliver a set FiO2. The accumulator also serves as a mufer for the blower.
The gas passes through the internal ow transducer, which measures ow, and then moves through the blower and
through the bacterial lter. Gas ows through the connected patient circuit to deliver the congured Ventilation therapy.
Patient Circuit
(pick one)
Internal
Bacterial
Filter
External
Bacterial
Filter
Internal Flow
Transducer
Inlet
Accumulator
Air Inlet Filter
FiO2
Sensor
Low-
Pressure
O2 Inlet
Ambient
Pressure
Transducer
Humidifier
Bypass
(optional)
Humidifier
(not provided)
Passive Patient Circuit with
Fixed Leak
Mouthpiece Patient Circuit
Valveless Patient Circuit
Vented Mask
(not provided)
Patient
Interface
(not
provided)
V*Home V
e
n
tilation
Th
erapy
Blower