Technical Specications
VentecLife.com 154
AC power
Sine AC mains waveform
Rated Voltage: 100 to 240 VAC, 50 to 60 Hz
Power for V+O+C+S+N(+Pro) congurations: ≤288 VA average, 420 VA peak
Power for V+C(+Pro) congurations: ≤180 VA average, 240 VA peak
DC power 20.0 to 30.3 VDC, as measured at the connection to the device, through 0 to 350 W instantaneous
loads
Power for V+O+C+S+N(+Pro) congurations: ≤200 W average, 350 W peak
Power for V+C(+Pro) congurations: ≤150 W average, 200 W peak
Intended for connection to 24 V (27.2 VDC typical) batteries
Female XLR3 connection port with the following pin conguration:
Pin 1 Signal: V+
Pin 2 Signal: RTN
Pin 3 Signal: N/A (Unused by V*Home)
Removable, Rechargeable Batteries
Category Specication
Operating time With ventilation at nominal settings and new batteries, approximate run time is 6 hours on set of
removable batteries.
Capacity 5.8 AHr
Voltage 14.4 VDC
Charging Each removable battery requires 5 hours, typical, from full discharge to full charge. If fully discharged,
charging time for the 2 removable batteries and Internal battery may require up to 14 hours.
Internal Rechargeable Battery
Category Specication
Operating time With ventilation at nominal settings and a new battery, approximate run time is 2 hours.
Capacity 3.9 AHr
Voltage 14.4 VDC
Charging If fully discharged, charging time for the 2 removable batteries and Internal battery may require up to
14 hours.
NOTE: The power supply provided with V*Home is >14 V. Charging time increases if a <14 V external
power source is applied.
12
3