Alarms
VentecLife.com 90
Alarm Settings Priority Description Recommended Action
System Fault
(all conditions)
N/A High The System Fault alarm activates if V*Home detects
any one of multiple system fault conditions.
Use the Event Log to
determine the System
Fault number, and then see
“System Fault Detection
Criteria and Recommended
Action” on page 177 and
take the corresponding
action.
Very Low FiO2
N/A High The Very Low FiO2 alarm activates when the
monitored FiO2 is less than 18%.
NOTE: If the FiO2 monitor is inactive, the Very
Low FiO2 alarm will be disabled. See “FiO2
Monitor” on page 99 for more information.
Remove any gas source
connected to V*Home and
monitor the patient.