33
Verathon Service Partner Manual: Repair & Replacement
REMOVE THE OVERMOLd ANd PCB
6. Using a Phillips bit, remove the (10) screws securing the overmold in place, and then remove the
overmold. This exposes the PCB assembly.
7. Carefully disconnect the three printer cables. Ensure that you do not damage, break, or bend the pins on
the PCB.
8. If there is a ferrite suppressor around the printer cables, remove it.