Visionix EyeRefract User Guide
89
Appendices
Immunity
test
Test level IEC
60601
Compliance
level
Electromagnetic environment –
guidance
Conduced
RF
IEC 61000-
4-6
Radiated RF
IEC 61000-
4-3
3V, 6V
10V/m
10V /m from 0.15
to 80MHz;
10V /m from 0.15
to 80MHz and
80% AM at 1kHz
10V/ m from
80MHz to 2.7GHz
3V, 6V
10V /m
10V/ m from
0.15 to 80MHz;
10V/ m from
0.15 to 80MHz
and 80% AM at
1kHz
10V /m from
80MHz to
2.7GHz
Portable and mobile RF
communications equipment should
be used no closer to any part of the
EyeRefract, including clables, than
the recommended separation
distance calculated from the
equation applicable to the frequency
of the transmitter.
Recommended separation
distance
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
manufacturer and d is the
recommended separation distance
in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey a,
should be less than the compliance
level in each frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol: