300004969171/B Page 111 of 148
Chapter 17: Maintenance
Service
The Volcano system contains no user-serviceable components. To avoid electrical shock,
do not remove any panels or covers. In the event of malfunction or system damage, turn
the system off, unplug the system from the power receptacle, and contact a Volcano-
certified service person and/or Volcano Technical Support.
As part of the hospital’s maintenance program, routine safety inspections and testing
should be conducted annually. Trained hospital technicians should inspect and safety test
all isolated connections and the system’s power supply.
CAUTION: The Volcano system must be serviced only by Volcano Corporation
Service personnel or by Volcano-certified technical representative.
Maintenance Frequency
User-Performed Maintenance Frequency
Archive patient cases Daily as used
Clean components in patient vicinity Daily as used
Clean fan vent and back of PC
PIM cable inspection and placement Daily as used
Verify cable connections Monthly
Inspect keyboard slides Monthly
Inspect and safety test all isolated
connections and system power supply
Annually
Volcano-Certified Maintenance Frequency
A comprehensive preventive maintenance
procedure performed by Volcano-certified
service engineers only
Annually, starting 18
months post-installation
NOTE: As part of the hospital’s maintenance program, routine safety inspections and
testing should be performed annually. Trained hospital technicians should inspect and
safety-test all isolated connections.