Chapter 16: Troubleshooting
Message Alerts
Upon startup the BIOS battery is failing.
representative to replace the
BIOS battery. Note: Until
battery is replaced, current
day/date will need to be
manually reset each instance of
power on.
The patient interface module does not have
a catheter plugged in.
The catheter is drawing too much current
from the patient interface module.
message>
The catheter produces a blank (dark) image
in the system.
Replace the catheter. If the
condition persists, call
Technical Support.
PIM Power Failure
The PIM has detected an undesirable high
power condition for the catheter in use. The
message disappears approximately two
seconds after the catheter has been
unplugged from the PIM, or when the high
power condition is corrected.
Replace the catheter. If the
message persists, call
Technical Support.
PIM not Detected
The system is unable to communicate with
the PIM unit.
If the message persists, call
Technical Support.
No Paper or Ink
Printer is out of paper or ink. Replenish the paper or ink
PIM Configuration in
progress. Please wait.
The hardware must be configured whenever
a PIM is changed. This message will
disappear once configuration is complete.
Wait for the message to
disappear. The wait time of this
message is between a few
seconds to two or three
minutes.
System error detected.
Auto-repair in progress,
please wait.
When the hardware experiences a failure,
the system will try to repair itself by
resetting either the VH or IVUS card.
Wait for the message to
disappear. If reset fails, the
system will shutdown.