EasyManuals Logo

Volcano CORE Mobile User Manual

Volcano CORE Mobile
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
300004969171/B Page 45 of 148
NOTE: Leave the power on during the case unless there is an emergency.
Connect the PIM
The IVUS catheter connects to the system via the patient interface module (PIM). The
PIM is typically left connected to the Connection Panel between cases, but may need to
be reconnected if removed after the prior case.
Notification of PIM connection status is provided in the lower right hand corner of the
screen. The lit green lamp confirms that a PIM is connected. In this case, the name of
the PIM (Phased Array or Rotational) is displayed depending on which PIM is connected.
To connect the PIM:
1 Connect the CPU end of the PIM cable into the PIM connector of the rear CPU
panel. Lock it in place using the retaining clips.
2 Connect the patient side of the PIM cable to the rear of the PIM. Orient the marker
on the cable connector to align with the indicator on the PIM, and slide the
connector in. If you need to remove the cable from the PIM, rotate the collar on
the PIM cable connector counterclockwise a quarter turn and then pull back.
CAUTION: Use care when handling the PIM to avoid accidental dropping, especially
when a catheter is connected.
NOTE 1: Leave the power on during the case unless there is an emergency.
NOTE 2: If the Volcano logo indicator is not lit, make sure that the system power is on
and plugged into a hospital grade grounded outlet. Make sure the fuse switch is turned
on (I symbol).
NOTE 3: The equipotentiality connector is provided on back panel of mobile chassis and
should be connected to the equipotentiality connector at the patient table.
(Equipotentiality cords are available from Volcano Service: PN 804768001 singled
ended, PN 804769001 double ended.)
Lock the rolling casters
It is important to lock the two casters in the front of the system to prevent movement
during the procedure. To lock the casters, press down on the tab with your foot. To
unlock the castors, use your foot to lift the tab up.
Prepare the Catheter
The IVUS catheter should be selected based on the needs of the procedure. Proper
handling and preparation of the catheter is critical to successful imaging. Prepare the
catheter in accordance with the Instructions for Use (IFU) included with each catheter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Volcano CORE Mobile and is the answer not in the manual?

Volcano CORE Mobile Specifications

General IconGeneral
BrandVolcano
ModelCORE Mobile
CategoryMouse
LanguageEnglish