Table 3 summarizes the sensitivity, specificity, and predictive accuracy results.
Table 3. The predictive accuracy, sensitivity, and specificity of the Volcano system
Eagle Eye VH classification tree by tissue type.
Tissue Type
Number of
Evaluation
Regions
Predictive
Accuracy (%)
Sensitivity (%) Specificity (%)
Fibrous Tissue
471
93.5
95.7
90.9
Fibro-Fatty 130 94.1 72.3 97.9
Necrotic Core 132 95.8 91.7 96.6
Dense Calcium 156 96.7 86.5 98.9
References:
1) Nair, A., M. P. Margolis, et al. (. "Automated coronary plaque characterisation with
intravascular ultrasound backscatter: ex vivo validation." EuroIntervention 3(9): 113-120.