Page 100 of 148 300004969171/B
to execute the Health
Checker module. Please
contact Volcano
Technical Support to fix
the problem. Click
'Cancel' to shut down the
system
(RECOMMENDED).
Click 'OK' to continue
with system start-up.
On startup, the health checker determines
the state of the hardware. When cards fail,
this message appears.
down. The power
disconnected from the system
for 10 seconds. Then start up
system. If startup fails again,
call Technical Support.
The current case is still
open. Please END the
current case before
switching to FFR Mode.
The system cannot switch to FFR if there is
an open case.
End or delete the open case
before switching to FFR.
The Meridian could not
stop the VH+ and IVUS
boards to switch to FFR
Mode.
Failure stopping either the VH+ board or
IVUS board while attempting to switch to
FFR mode.
End/archive the current IVUS
case, shutdown the system.
Wait for
1 minute before
powering up again.
restart the VH+ and
IVUS boards back to
Meridian Mode.
Failure to start either the VH+ or IVUS
board after switching back from FFR to
IVUS mode.
Shutdown the system. Wait for
1 minute before powering up
again.
before ending case.
Please end or delete the
open case.
The system was shutdown prematurely
before a case was ended. On startup, the
system goes in recover mode and warns the
user to end or delete the open case.
End or delete the open case.
system will abort the case
archiving.
Pressing the power button while archiving.
Selecting yes will turn off the
application.
An error occurred
during initialization.
Please shut down and
then restart the system.
Contact Volcano
Technical Support if
issues persist.
The Health Checker encountered an
unrecoverable error on startup.
Shutdown the system. Wait for
1 minute before powering up
again. If the problem persists,
call Technical Support.
Error Opening
Association: <variable
text>.
There is a problem with the worklist
network connection.
Check connection to the
worklist server