EasyManuals Logo

Volcano CORE Mobile User Manual

Volcano CORE Mobile
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
Page 120 of 148 300004969171/B
Power SpecificationsInput
Maximum Continuous
Power
550 watts
Hold-up Time
>40ms @ full load
Over voltage Protection
+15% of set voltage
Overload Protection
Auto Recover
Minimum Loading
None
Cross regulation
<0.5%
Line & Load regulation
<0.5%
Turn on Time
< 2 sec
EMC Compliance
Compliance to the following standards:
EN61000-4-2 ESD level3
EN61000-4-3 RF susceptibility 3v/m
EN61000-4-4 EFT level3
EN61000-4-5 Surge level3
EN61000-3-2 Harmonics
EN61000-3-3 Flicker
EN61000-4-11 Voltage dips, short interruptions
EN55011 conducted an radiated emissions
Fusing
The Isolation Transformer is merged with a fuse on each mains leg. The fuse values are
dependent on the transformer setting for the available AC mains power from the wall.
The transformer uses 5x20mm time delay fuses.
The AC mains input for the workstation power supply and 48VDC power supply are
internally fused and are not serviceable parts.
The 48VDC power supply output is protected with a 5A, 5x20mm time delay fuse.