Recommended Preparation Procedures 55
Figure 6-6 outlines how to change from a buffered solvent to a totally organic solvent in a
degasser-based Separations Module.
Note:
All flow chart references refer to sections or figures in the Waters 2690 Separations
Module Operator’s Guide.
Figure 6-6 Changing from a Buffered Solvent to an Organic Solvent
(Degasser-Based Separations Module)
Remove the Solvent
Line from the
Reservoir with the
Buffered Solvent
Purge the Injector
for a Minimum of
6 Loop Volumes
(See “Purging the
Injector” in
Section 4.4.2)
Using a Flow Rate of
2.000
mL/min,
Perform a Wet Prime
Until Air Begins to
Enter the Line
(See Section 4.4.2)
Fill a Clean Solvent
Reservoir with the
Organic Solvent and
Place the Solvent
Line in
the Reservoir
Start
Place the Previously
Removed Solvent
Line into Filtered
HPLC-Grade Water
and Perform a Wet
Prime for
10
min at
5
mL/min
(See Section 4.4.2)
Perform the Steps
in
Figure 6-2
Remove the Solvent
Line from the
HPLC-Grade Water
Equilibrate
a
the
Solvent in the
Degasser Chamber
by Setting a Flow
Rate of
0.000
mL/min for
5
min
(See “Equilibrating
Solvents in the
Vacuum Degasser”
in Section 4.4.1)
a
Except for the short period of time required to equilibrate solvents in the degasser, do not run the degasser
STOP
Attention:
To avoid having salts precipitate in the Separations Module, use an
intermediate solvent such as water when you change from buffers to
high-organic-content solvents. Refer to Section D.3, Solvent Miscibility, for
information on solvent miscibility.