November 16, 2016, 715004521 Rev. C
Page 74
Figure 5–29: Core 4 configuration with a PDA, ISM, and QDa
The following table describes final tubing connections and does not list factory-installed tubing
connections.
Table 5–14: Core 4 configuration with a PDA, ISM, and QDa
Tube
Tube
description
From To ID
Notes:
recommended
trim length
1 60.0 inch
PEEK tubing
ISM splitter
module
ccManager vent
valve tee
0.007 inches Optional trim up
to 30 inches
2 28.0 inch
PEEK tubing
CM-30S or PDA ISM splitter
module
0.007 inches Trim to
15 inches
3 Probe ISM splitter
module
QDa probe N/A N/A
Solvent tray
ccManager
SM-FL
ccBSM
CM-30S
QDa
ISM with
UPC
2
splitter
PDA
1
2
3