< 5. Key Operation and Display >
5-12
IM 12F5A1-01E
(a)<*OUTPU>(Outputrangesettingmode)
Press the
key to enter setting level and ,
ENT
to select <*OUTPU>.
Set concentration value corresponding to each output span.Set span to at least 1.00 mg/L (or ppm).
When service code 05 “Two-line-segment output” is set “ON” then set 50% and 100% outputs. Default
for this parameter is “OFF” .
*
OUTPU
NO
<
*
SET.HD>
ENT
ENT
*
50.0% 1
1 . 0 0
*
100.0% 1
2 . 0 0
YES
F5-12-1e.ai
Range 1, output 50% value setting
(applies only Two-line-segment output)
Range 1, output 100% value setting
Range 2, output 50% value setting
(applies only when Two-line-segment output is ON
and remote range switching is ON.)
*
50.0% 2
1 . 5 0
*
100.0% 2
3 . 0 0
ENT
ENT
Range 2, output 100% value setting
(When remote range switching is ON.)
(b)<*SET.HD>(HoldParameterSettingMode)
Press the
key to enter the setting level and press the
NO
keytoselect<*SET.HD>.
Pressing the
YES
keywith“*HLD.ON”displayedcausestheFC400Gtoentercalibration,setting
level, and service level, automatically holding output. Pressing the
YES
keywith“*HLD.OF”
displayed cancels holding of output.
When“*HLD.ON”isselected,selectthevalueimmediatelyprecedinghold(press
YES
key with
“*HD.LST”displayed)orxedvalue(press
YES
keywith“*HD.FIX”displayedandmakesettingwith
“*HOLD.%”displayed)asaholdoutputvalue.
To cancel hold, use the hold ON / OFF mode in the operation level.
*
SET.HD
NO
<
*
WASH>
*
HOLD.%
1 2 . 0
YES
F5.12.2e.ai
*
HLD.ON
*
HD.LST
*
HLD.OF
*
HD.FIX
YES
YES
NO
NO
NO
NO
YES YES
ENT
(Note)
To set hold
(Note)
Previous value
To cancel hold
Fixed value
Enter the fixed value in hold status.
Note : The status selected
previously appears first.