< 9. Troubleshooting >
9-4
IM 12F5A1-01E
9.2 FortheCaseWhereNoErrorIndicationAppears
Therearecaseswherenormalmeasurementisprevented,suchasinuctuatingsetvalueeven
though no error is detected. Table 9.2 shows such abnormalities, possible causes, and remedies.
Table9.2 Abnormalities,TheirPossibleCauses,andRemedies
Abnormalities PossibleCauses Remedies
Measured value is
abnormal.
1. Indicator electrode has stopped.
a. Electrode drive belt does not turn.
b. Cell motor does not run.
2. Indicator electrode has worn out.
3.Insucientcontinuityattheelectrode
mechanism block.
1. Take steps as follows :
a.Sucientlytightenthesetscrew.
b. Replace the motor with a new one.
2. Replace the electrode with a new one.
3. Replace the rotating contact (see Sec.
8.2.1).
Measurement error
is large.
1. Indicator electrode has been
contaminated.
2.Zeroand/orspanhasnotbeen
correctlyadjusted.
3. Auto zero calibration abnormal.
1. Polish the indicator electrode in
accordancewithSec.8.1.1also,clean
the ceramic beads in accordance
withSec.8.1.2(ifitisnotpossibleto
sucientlycleanthebeads,replace
them with new ones).
2. Conduct zero and / or span calibration.
3a.Replaceactivatedcharcoallter
3b. Piping related to auto zero calibration
is blocked
Measured value
uctuateslargely.
1. Indicator electrode does not rotate at
constant speed.
2.Insucientcontinuityattheelectrode
mechanism block
1. Check if the set screws have loosened.
Check if the belt, driven shaft assembly
is normal.
If not, replace them.
2. Replace the rotating contact.
Measured value
does not vary.
1. Measuring tank has no sample
water.
2. Electrode to electrode circuit is open.
3. No voltage is applied between the
indicator and counter electrodes.
1. Repair abnormality such as piping
disconnection or sample water feed
failure to normal condition
2. Repair open circuit. If open circuit
occurs in an electrode, replace such
electrode.
3. Check the converter.
Abnormal analog
output
1. Lead resistance is too large.
2. Output circuit abnormal.
1.Reducetheresistanceto550Ωorless.
2. Contact YOKOGAWA.
The reading
deviatessignicantly
in the positive
(or negative)
direction, or
overshoot occurs.
1. Span calibration has not been
performed correctly. Span calibration
was performed with electrode
sensitivity initialization set to ON (1)
inservicecode08("CHANG"Mode)
and using a calibration solution with
alargedierenceinconcentration
from the input concentration value.
1. Set electrode sensitivity initialization to
ON(1)inservicecode08("CHANG"
Mode) and perform calibration again.
(See item (f) in Sec. 5.4.2, Sec. 7.3)