EasyManuals Logo

YOKOGAWA FC400G User Manual

YOKOGAWA FC400G
91 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
< 7. Calibration >
7-3
IM 12F5A1-01E
7.2.2 MethodofZeroCalibrationusingChlorine-freeWater
ThemethodofperformingZeroCalibrationusingChlorine-freeWaterisasfollows:
Note that the electrode mechanism block should be secured to the measuring tank, and the
electrodes rotating.
(1) Prepare 2-3 liters of chlorine-free water.
Thechlorine-freewatershouldconsistofpurewatertowhich10mlofpH7bu󰀨ersolutionand
50 mg of NaCl has been added.
(2) Select<CAL>mode(“ZERO”displayedinmessagedisplay.)
Note: IfServiceLevelcode08issetto“1”for“initializeelectrodesensitivity”,thenplacingtheunitin<CAL>moderesultsin
“NO.CHNG”or“CHANGE”beingdisplayed.Inresponse(seeSec.7.3)pressthe
YES
keyand“ZERO”isdisplayed.
(3) Connect piping to supply chlorine-free water to the FC400G.
Close the needle valve at the sample water inlet of the FC400G, and remove the sample
waterpipingfromthejointtothemeasuringtank.Releaseanysamplewaterremaininginthe
measuring tank by opening the drain cock on the tank. Use tubing and hardware to connect the
chlorine-freewatertanktothesamplewaterinlet.Useaheaddi󰀨erenceorapumptosupply
chlorine-freewaterataowrateofatleast100ml/min.
(4) Withthechlorine-freewaterowingatthespeciedowrate,watchthedatadisplayandwaitfor
it to stabilize.
(5) Perform zero calibration
Check that the data displayed on the data display has stabilized, then press the
YES
key.
While calibration is being performed, a blinking “mg/l” or “ppm” is displayed. When calibration
endsnormally,“ZR.END”isdisplayed.
Note: Duringcalibration,ifanabnormalityoccursthen“END.E6”or“END.E7”aredisplayed.RefertoSec.7.4foradescription
ofhowtoxtheproblem.
(6) ExitingZerocalibration
Conrmthatanormal“ZR.END”ratherthananerrormessagethenpressthe
YES
key.
Pressing the
YES
key switches the display to “SPAN”. After stopping the indicator electrode,
reverttheelectrodetoitspositioninthemeasuringtank.Afterxingtheelectrodemechanism
block to the measuring tank, restart electrode rotation. Perform span calibration (see Sec. 7.3)
after performing zero calibration.
7.2.3 AutoZeroCalibrationusingChlorine-free (chlorinenotincluded)Water.
AutoZeroCalibrationisperformedusinganactivatedcharcoallter.Setupasfollows:
(1) In Service Level code 09 [Washing / Auto zero calibration] set “1” for Auto zero calibration.
The default is “0” for washing.
(2) Select“*AZ.CAL”forAutozerocalibrationsequenceinSettingLevel.
Afterselecting“AutoZeroCalibrationtimerON(*T.ON),selectCalibrationInterval(*INTD),
Calibration Setup time (WT.min), and Stabilization (settling) time (*RT.min).
Defaultvaluesare:ForCalibrationInterval(*INTD),10days(refertozerolterushing
functions); for Calibration Setup time (WT. min), 10 min ., for Stabilization (settling) time (*RT.
min), 10 min.
WhenAutoZeroCalibrationstarts,theanalogoutputisheld,<HOLD>lights,and<CAL>(blinking)is
displayed. At the same time, the <MAINT> contact output turns ON.
After the calibration setup time has elapsed, the measured value drops to near zero, and after this
measured value has stabilized it is automatically reset to zero. The method used to determine if the
value has stabilized is the same as for manual calibration.
Whencalibrationiscompleted,theowthroughtheunitisswitchedfromzerowater(puredeionized
water) to sample water, then after the stabilization (settling) time has elapsed the analog output hold is
released,<HOLD>and<CAL>goOFF,andthe<MAINT>contactturnsOFF.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA FC400G and is the answer not in the manual?

YOKOGAWA FC400G Specifications

General IconGeneral
BrandYOKOGAWA
ModelFC400G
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals