EasyManuals Logo

YOKOGAWA FC400G User Manual

YOKOGAWA FC400G
91 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
< 7. Calibration >
7-1
IM 12F5A1-01E
7. CALIBRATION
Calibration is required to correctly measure free chlorine concentrations using the FC400G.
Calibration is performed for the zero point and span. There are two methods to calibrate the zero
point: open input circuit and chlorine-free-water-based measurement methods. Also, there are two
span calibration methods: sampling and standard solution methods. Calibration is performed at start-
up or when operation is restarted after being stopped for a long period of time.
Also, calibration is performed periodically (recommended every month or so) during operation. This
chapter describes the calibration procedure.
7.1 Running-in
In general, calibration requires running-in. Always run in the FC400G at start-up or when it restarts
after being temporarily stopped.
Running-inrequires,rst,polishingoftheindicatorelectrode.ComplywithSec.6.1.4onhowtopolish
the indicator electrode.
After polishing, install the polished indicator electrode and select “CELL. ON” in the <CELL> mode.
Then press the
YES
key to rotate the electrode. Operate the free available chlorine analyzer for
approx.onehour.
Calibrationcanbeconductedafterthisapprox.1hrrunning-in.
Note: It is recommended that electrode polishing be performed even when calibration is performed periodically during operation.
However,electrodepolishingcanbeomittedifthecalibrationintervalisshortanditcanbedeterminedthattheelectrodeisnot
contaminated.
The characteristic readings of the indicator electrode are temporarily not stable immediately after polishing the electrode.
Therefore,afterelectrodepolishingisperformed,leavetheFC400Grunningforawhile(approx.1hour)underthemeasurement
conditions until the characteristic readings of the indicator electrode stabilize. Calibration is performed following this running-in.
7.2 ZeroCalibration
The method of calibrating the zero point of the free available chlorine analyzer includes two
approaches; open input circuit and chlorine-free-water-based measurement methods.
Intheformertheelectrodeisexposedintheairsothatnocurrentowsbetweentheindicator
electrode (rotating electrode) and counter electrode. The latter uses chlorine-free water (e.g., water
lteredthroughactivatedcharcoal).
Duringautozerocalibrationwiththeactivatedcharcoallter,chlorine-free(chlorinenotincluded)
water must be used.
If the FC400G generally measures free chlorine concentrations as low as 0.1 mg/l, conduct zero
calibration using the chlorine-free-water-based measuring method.
However,theopeninputcircuitmethod,easytoconduct,isgenerallyused.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA FC400G and is the answer not in the manual?

YOKOGAWA FC400G Specifications

General IconGeneral
BrandYOKOGAWA
ModelFC400G
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals