< 6. Operation >
6-8
IM 12F5A1-01E
6.3.2 SettingintheSettingLevel
Unscrew four screws in the front panel of the FC400G converter to remove the converter case cover.
Then press the
key to select the setting level.
(a) Output range setting (<*OUTPU> Mode)
When straight-line output is selected using SERVICE code 05, you need to set only the
concentration value corresponding to output of 100% . The setting range is 0.00 to 3.00 mg/L,
but the span should be set to at least 1.00 mg/L.
When remote range switching is turned ON using SERVICE code 10, then set the concentration
values corresponding to outputs of 100% for both ranges 1 and 2.
When two-line-segment output is selected using SERVICE code 05, and remote range switching
is also turned ON using SERVICE code 10, then set the concentration values corresponding to
outputs of 50% and 100% for both ranges 1 and 2. Shipping-time default settings are:
Range1 50% : 1.00 mg/L, 100% : 2.00 mg/L
Range2 50% : 1.50 mg/L, 100% : 3.00 mg/L
(b) HoldParametersSetting (<*SET.HD>Mode)
Set various data for hold of output during maintenance.
• OutputHoldON/OFFSelection
Select*HLD.OFtodisableoutputholdor*HLD.ONtoenableoutputholdduringmaintenance.
•Selecting*HLD.ONcausesoutputtobeautomaticallyheldduringcalibrationofzeropointand/
or span, automatic washing, and entry into the setting level by pressing the
key.
Further,outputholdcanbemanuallymadeinthe<HOLD>modeintheoperationlevel.When
the FC400G is in hold status, appears on the LCD.
Note:Selectionof*HLD.OFinthesettingleveldisablesentryintothe<HOLD>modeintheoperationlevel.
•HoldOutputValueSelection(when*HLD.ONisselected)
Select*HD.LSTtosetthevalueimmediatelyprecedingholdasholdoutputvalueor*HD.FIXto
setaxedoutputvalue(presetvalue).
•SettingthexedOutputValue(when*HD.FIXisselected)
Setxedoutputvalueintherangeof0.0to100.0%.Atfactoryshipment,thexedoutputvalue
is set to 50.0%.
(c) Wash / Auto zero calibration parameter setting (<*WASH>,<*AZ.CAL>Mode)
Washparametersetting<*WASH>Mode
This mode is not generally used since the standard free available chlorine analyzer has no
washing function. Use this mode only if the washing function is added to the FC400G for special
specications.Therefore,*T.OFFmustgenerallybeestablished.
SettingforSpecialSpecicationoftheFC400Gisasfollows.
• Washing Parameter Setting / Manual Washing Selection
To set up various parameters for washing timer, select *TIMER. For manual Washing, select
*START .
Press the
YES
key. This starts washing and message display becomes “STOP”. To stop
washing, re-press the
YES
key.
• Washing Timer ON/ OFF Selection
If the washing timer is not used (for the standard free available chlorine analyzer), always select
*T. OFF. Select *T. ON only when the washing timer is used.
• Washing Interval Setting (when *T. ON is selected)
Setwashinginterval*INTHintherangeof0.1to24.0hrs.Thewashingintervalissetto2.0hrs
at factory shipment.