< 6. Operation >
6-7
IM 12F5A1-01E
Table6.2 SettingItemList(SetValuesatShipmentandTheirSettingatRange)-2
Setting Item
InitialValue
(at Factory Shipment)
Selection and Setting Range
Service
Level
Code 01 (*TEMP) Mode
•Temperature unit selection °C ( 0 ) °C ( 0 ), °F ( 1 )
Code 02 (*FAIL) Mode
•Reserved code 0 ( 0 ), ( 1 )
Successive operation is not
aectedwhether(0)or(1)is
selected.
•Selection of temperature compensation
rangeexceedingfailcontactoutputON/OFF
OFF ( 0 ) OFF ( 0 ), ON ( 1 )
Code04(*∆T.SEC) Mode
•Setting of stability determination parameter
*∆T.SEC
10 [ sec ] 5 to 200 [ sec ]
•Setting of stability determination parameter
*∆PPM
0.01 [ mg/l ] 0.01 to 0.05 [ mg/l ]
•Setting of stability determination parameter
*CT.min
7.0 [ min ] 0.1 to 25.0 [ min ]
Code 05 (*NONLN) Mode
•Selection of linear output / two - line- segment
output
Liner output ( 0 ) Linear output ( 0 ),
Two-line-segment output ( 1 )
Code 07 (*DRV V) Mode
•Setting of applied voltage compensation
parameter Start, *DRV V
-0.10 [ V ] (Note 1) -1.50 to 1.50 [ V ]
•Setting of applied voltage compensation
parameterSlope,*V/μA
-0.02[V/μA](Note2) -0.10to0.00[V/μA]
Code08(*CHANG) Mode
•Selection of electrode sensitivity initialization
ON /OFF
OFF ( 0 ) OFF ( 0 ), ON ( 1 )
Code 09 (*WA / CA) Mode
*Wash-AutoZeroCalibration Wash ( 0 ) (Note 3) Wash ( 0 ),
AutoZeroCalibration(1)
Code 10 (*R.RNGE) Mode
*Remote Range switching ON / OFF selection OFF ( 0 ) (Note 4) OFF ( 0 ), ON ( 1 )
(Note1)Setto+0.1Vin(todecreaseinuenceof)combinedchlorineinsensitiveversion
(Note2)Setto-0.01V/µAin(todecreaseinuenceof)combinedchlorineinsensitiveversion
(Note3)WhenST401G/AZCisspecied,initialvalueisAutoZeroCalibration(1).
(Note4)WhenST401G/RRisspecied,initialvalueisON(1).