EasyManuals Logo

YOKOGAWA FC400G User Manual

YOKOGAWA FC400G
91 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
< 6. Operation >
6-11
IM 12F5A1-01E
6.3.3 SettingintheServiceLevel
Press the
YES
key to enter the service level on *SERVC display in Setting Level.
Enter a code in the service level to enter the service mode.
(1) Code 01: Selecting Temperature Unit (°C / °F) (*TEMP Mode)
Set temperature unit to “0” to apply °C or “1” to use °F.
This parameter is set to “0” (°C) at factory shipment.
(2) Code 02: SelectingExceededTemperatureCompensationRangeCheckON/OFF(*FAIL
Mode)
Note:Theleftdigitoftwo-digitguresisthereservedcode.Thisdoesnota󰀨ectoperationwhetheritis0or1.
To check if sample water temperature is in the temperature compensation range (0 to 40°C), set
therightdigitoftheguresto“1”.Setitto“0”whennotconductingacheck.
When this parameter is set to “ 1 ” (conducting the check), fail contact output becomes ON if
samplewatertemperatureexceedsthetemperaturecompensationrange.
This parameter is set to “ 0 ” (no check) at factory shipment.
(3) Code 04: SettingStabilityDeterminationParameters(*∆T.SECMode)
Set parameters (PPM/∆TandCT)usedtodetermineifelectrodeoperationisstableduring
zero point and / or span calibration.
If concentration changes by PPMmg/lormoreduring∆Tsecondsevenwhenthetimeelapses
for CT minutes, meaning that operation is unstable, error indication is generated,
•Setting*∆T.SEC(StabilityCheckTime)
Setthestabilitychecktime(*∆T.SEC)intherangeof5to200sec.Thisparameterissetto10
sec at factory shipment.
• Setting *PPM (Concentration Change Allowance)
Set the concentration change allowance (*PPM) in the range of 0.01 to 0.05 mg/l. This
parameter is set to 0.01 mg/l at factory shipment.
• Setting *CT. min (Stabilization Wait Time)
Set stabilization wait time (*CT.min) in the range of 0.1 to 25.0 min. This parameter is set to 7.0
min at factory shipment.
(4) Code 05: Selecting Linear Output / Two-line-segment Output (*NONLN Mode)
Settheparameterto“0”togeneratelinearoutput(seeFigure6.6)intherangeof0to100%or
to “ 1 ” to generate two-line-segment output where output is segmented into two steps; 0 to 50%
and 50 to 100%. This parameter is set to “0” (linear output) at factory shipment.
100
50
0
1 20
100
50
0
10.5 20
F6.6e.ai
Output
(%)
Output range (mg/l)
Output
(%)
Output range (mg/l)
Figure6.6LinearOutputExample Figure6.7Two-Line-SegmentOutputExample

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA FC400G and is the answer not in the manual?

YOKOGAWA FC400G Specifications

General IconGeneral
BrandYOKOGAWA
ModelFC400G
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals