< 9. Troubleshooting >
9-3
IM 12F5A1-01E
Err.10:(Converterfailure)
The FC400G free available chlorine analyzer has become defective. Contact Yokogawa service.
NOTE.01:(InputDataSettingRangeExceeded)
Thiserrorappearsifasetvalueexceedstherelevantsettingrangewhendataissetintheoperation
level, setting level, or service level.
If this error message appears, the relevant data will not be updated. Set correct value again.
NOTE.02:(InputDataSettingError)
This error is generated if the mg/l (or ppm) value for 100% signal output falls below 1.00 mg/l (or 1.00
ppm) in the <*OUTPU> mode in the setting level.
If this error message appears, the relevant data will not be updated. Set correct value again.
NOTE.03:(InputDataSettingError)
This error is generated if the mg/l (or ppm) value for 100% signal output is set to a value below the
value for 50% signal output when two-line-segment output is set in service code 05 in the <*OUTPU>
mode in the setting level.
If this error message appears, the relevant data will not be updated. Set correct value again so that
themg/l(orppm)valuefor100%signaloutputexceedsthevaluefor50%signaloutput.