EasyManua.ls Logo

YOKOGAWA FC400G

YOKOGAWA FC400G
91 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
< 9. Troubleshooting >
9-2
IM 12F5A1-01E
Err.1:(Di󰀨usionCurrentValueExceeded)
“Err.1”appearsifthedi󰀨usioncurrentduringmeasurementexceedstherangeof-9.00to26.00µA.
Conrmiftheappliedvoltageparametervaluesetinservicecode07iscorrect.Also,conrmthat
there is no grime or moisture adhesion resulting in insulation deterioration on the electrode cable
connection.
Err.2:(SampleTemperatureError)
“Err.2”appearsifsampletemperatureexceedstherangeof0.0to50.0°C.
Control sample temperature to be in the range of 0.0 to 50.0°C.
Iftheerrorindicationfailstoextinguishevenwhenthesampletemperatureisintherangeof0.0to
50.0 °C, RTD is possibly defective (open circuited). In this case, replace the counter electrode.
Err.4:(AppliedVoltageError)
“Err.4”appearsiftheappliedvoltageoutputexceedstherangeof-1.5to1.5V.
Conrmiftheappliedvoltageparametervaluesetinservicecode07iscorrect.
Err.5:(SlopeError)
“Err. 5 appears if the initial sensitivity value ( value set to initialize electrode sensitivity in service
code08)exceedstherangeof50to120%whentheslopeisconrmedduringcalculationinspan
calibration.
The calibration standard solution used for span calibration may have aged, entering the incorrect
concentration value, or the indicator electrode may have deteriorated.
Conrmthepresenceofanyofthese.
Iftheerrorstillfailstoextinguishevenwhenre-calibrationismadeusinganewcalibrationstandard
solution or after the indicator electrode has been polished using alumina, the indicator electrode must
be replaced.
Err.6:(ZeroPointError)
“Err.6”appearsifthezeropointexceeds±1.00µAwhenthezeropointisconrmedduringcalculation
in zero calibration.
Chlorine may enter the chlorine-free water used for the calibration (when the chlorine-free-water-
based measuring method applies), or the indicator electrode or counter electrode may be defective.
Conrmifanyoftheseisthecause.
Err.7:(ResponseErrorduringZeroand/orSpanCalibration)
“Err. 7” appears if concentration does not stabilize within “*CT. min” (stabilization setup time) set in
service code 04 during zero and / or span calibration.
Conrmresponserecoveryaftertheindicatorelectrodehasbeenpolishedusingalumina.Also,
conrmthatthesetvaluesofstabilitydeterminationparameters“*ΔT.SEC”,“*ΔPPM”,and“*CT.min”
set in service code 04 are correct. If the parameters are correct, but still no response recovery, the
indicator electrode may have deteriorated.
Replace the indicator electrode.
Err.8:(TemperatureCompensationRangeExceeded)
“Err.8”appearsifsampletemperatureexceedsthetemperaturecompensationrangeof0.0to40.0°C.
Control sample temperature to be in the range of 0.0 to 40.0°C.
Note: ExceededtemperaturecompensationrangecheckisavailableonlywhentemperaturecompensationrangefailONisselectedin
service code 02.

Table of Contents

Related product manuals