< 5. Key Operation and Display >
5-14
IM 12F5A1-01E
(d)<*AZ.CAL>(AutoZeroCalibrationmode)
Theactivatedcharcoallterisusedandautocalibrationisperformed.ToturnthisfunctionON,service
code09[Wash/AutoCalibration]switchissetto“1”.Thedefaultsettingis“0”forWash.ForAutoZero
Calibration mode, the calibration interval “*INT D” (unit: days), calibration setup time “*WT.min” (unit:
minutes), and stabilization (settling) time “*RT.min” . (unit: minutes) are set. The optimum time settings
depend on the sample and the measurement environment.
*
INT D
10 . 0
F5.4.2-3a.ai
*
RT.min
0 5 0
*
T.ON
*
WT.min
0 1 0
*
T.OFF
TIMER OFF
TIMER ON
YES
ENT
ENT
ENT
YES
NO
NO
*
AZ.CAL
<
*
PLAT>
YES
Set calibration interval (Note 2)
Set calibration setup time
Set stabilization (setting) time
Note
Note 2 : When the calibration interval is more than
one day, the zero filter flushing functions
(to cleanse the activated charcoal filter and
help prevent poisoning of the filter) automatically
wash the filter with zero water once a day.
Note : The previously-selected status is displayed.
(e)<*PLAT>(PlateauMode)
Press the
key to enter the setting level and press the
key to select < *PLAT >.
Press the
YES
key with *MEAS displayed. When the plateau characteristic collection performs, enter
YES
key again.
In the plateau characteristic collection function (“ *MEAS”), setting initial applied voltage, step
applied-voltage and step count and then selecting “*START” causes mg/l (or ppm) indication to blink,
displayingtheappliedvoltage.Whenthedi󰀨usioncurrentvaluestabilizes,theappliedvoltageand
di󰀨usioncurrentvaluesarestoredinthememorytoapplythenextstepvoltage.Astepnotneeded
can be aborted using the
ENT
key. This erases data in that step (“---” appears when such data is
read),andtheoperationmovestothenextstep.
When the voltage has been applied to all steps, “*READ” appears. The data readout function
(“*READ”) displays in turn the data acquired previously each time the
YES
key is pressed.