<5. Others (Converter)>
5-7
IM 12A01G01-03EN 1st Edition : Mar. 25, 2021-00
5.1.2 Load conguration
You will load conguration data of all sensors connected to FLXA402T as current conguration
data. The screen of Load conguration shows present conguration set list. The gure below
shows an example.
Figure 5.7 Load conguration
When you tap one of the item (above), a dialog box appears.
Dialog box shows the conguration set you select. Here you cannot modify a conguration set le
name. Tap “Execute” to load the conguration le you select.
When the loading is completed successfully, FLXA402T returns to Home screen,
NOTE
If di󰀨erent sensor is connected for data loading from the one used for data saving, you cannot
load the data. An alarm dialog appears, because FLXA402T does not identify the sensor.
Example
Conguration set information was saved in FLXA402T when both sensor connection of 2-1
and 2-2 were connected.
If you try to load this conguration set while connecting only 2-2, an alarm dialog pops up,
because FLXA402T cannot load the conguration to 2-1, although it succeeds loading onto
2-2.
In this case, “Load individual conguration le” (5.1.1) would work for loading onto the sensor
which was not connected while FLXA402T was saving the data.