<5. Others (Converter)>
5-6
IM 12A01G01-03EN 1st Edition : Mar. 25, 2021-00
5.1.1 Load individual conguration File
Conguration les for FLXA402T converter or sensors can be loaded individually as current
value.
Go to Converter menu > Others > Save/Load > Load individ. cong. le
Select one below to see the list of conguration les.
Figure 5.4 Load Individ. cong. le
To load individual conguration le, tap one le in the list.
Figure 5.5 Load individ. cong. le
If you tap one of the les in the list above, a dialog box appears. (left below) In “File name”, you
will see the le name you selected. (left below) You cannot modify the le names.
On a dialog for loading sensor conguration le (right below), open the drop-down list next to
“Target sensor” to select one sensor connection. The drop-down list shows blank if there is no
valid sensor connection to load. “Execute” key does not work either.
Figure 5.6 sample (for converter) Sample (for pH)
Conrm the le you selected is displayed in the box of File name (left above). Tap “Execute.” The
le starts being loaded and conguration is modied.
When Converter conguration le is loaded successfully, FLXA402T returns to Home screen.
When sensor conguration le is successfully loaded, FLXA402T returns to the screen of Load
individ. cong. le, where you can select another le loading.