8.2.14 Simultaneous Ammonium and Potassium
This is a direct reading of Ammonium and Potassium in solution at the ISE.
Ammonium Potassium
Buffer
YSI 2970
Std
YSI 2972 YSI 2971
Std
YSI 7173 and YSI 7179
YSI 2974 YSI 2975
Probe
YSI 2976
Size
25 µL
Range
10–500
mg/L
20–1000
mg/L
Point
500 mg/L
1000
mg/L
Point
100 mg/L 200 mg/L
30 sec
Precision
(CV,n=10)
mg/L,
whichever
mg/L,
whichever
±5%
Working
Life
90 days
Note: See Appendix B – Concentration
Unit Conversion if concentration unit
conversion is required.
• Sodium may be an interferent to the measurement of
ammonium with some samples. The interference is
relatively small on a weight/weight basis (<1%). The
2900 Series does not automatically compensate for the
presence of sodium. In order to check the potential
sodium interference, YSI provides 7173 NaCl Check
Solution (4 g/L Na
+
). Presented as a sample, the effect
of sodium on the ammonium sensor (and to a lesser
extent on the potassium sensor) can be determined. If
you determine this sodium presence is a significant
factor in your measurement, you may use the result of
the 7173 measurement to correct ammonium readings.